Bristol wrenches

Print Friendly, PDF & Email
bristol-set-screw

bristol-set-screw

Most of the set screws used on older PHI optics are the Bristol set screw type. The main advantage to this drive system is that almost all of the turning force is applied at right angles to the fastener spline face, which reduces the possibility of stripping the fastener. For this reason Bristol set screw are often used in softer, non-ferrous metals. Compared to an Allen drive, Bristol set screws are less likely to strip for the same amount of torque. Bristol set screws come with 4 flutes on the smaller sizes and 6 flutes on larger sizes.

When working on optics such as ion sources, analyzers, specimen stages and so on it is not uncommon to strip the set screws when removing them. A trick that can help prevent striping is to use a small amount of isopropanol as a lubricant before removing the set screws.

Some sources for replacement Bristol set screws are –

Fasteners & Metal Products Corporation

Cam-Tech Industries

Radax Industries

Note that these set screws are not cleaned for UHV and so you will need to run them through a cleaning process before using them for UHV optics applications.

bristol-wrench

bristol-wrench

When using Bristol set screws you need to use Bristol spline drivers. A source for Bristol spline drivers is –

Bristol Wrench Company, Inc  4 Hershey Drive, Ansonia, CT  06401

http://www.bristolwrench.com/

Note that Bristol Wrench no longer sells individual spline wrenches, only in sets of 10.  Good news is that you can get individual Bristol spline wrenches from McMaster Carr –

https://www.mcmaster.com/bristol-splines/

Ion Pump Rejuvenation Procedure

Print Friendly, PDF & Email

After prolonged periods of sputtering with Argon gas, the ion pumps can become saturated, resulting in occasional “belches” of Argon during which the ion pumps overheat and release large amounts of gas. These belches usually result in a snowball effect that can dump the system. Rejuvenating the ion pumps once every few months (more often if you do a lot of sputtering) will help to prevent the belch problem from recurring.

To rejuvenate the ion pumps with O2:

1. Turn off all filaments, including the ionization tube (DIG).

2. Set the ion pump control panel meter to the 200mA current settings and set the

ion pump to the run (protected) mode.

3. Slowly bleed in O2 until there are 40mA of current shown on the ion pump panel meter. You will need to change ranges on the panel meter as the current is increased.

4. Adjust the leak valve as needed to maintain 40mA of current for 20 to 30 minutes.

5. Close the leak valve. It takes about one day for the vacuum to return to its previous level.

 

For more information on rebuilding ion pumps, search for Ion Pump in the RBD TechSpot blog search box.

For more information on ion pump theory, here is a link to an informative paper – https://cds.cern.ch/record/454179/files/p37.pdf

And, from Wikipedia:

An ion pump (also referred to as a sputter ion pump) is a type of vacuum pump capable of reaching pressures as low as 10−11 mbar under ideal conditions.An ion pump ionizes gas within the vessel it is attached to and employs a strong electrical potential, typically 3kV to 7kV, which allows the ions to accelerate into and be captured by a solid electrode and its residue.

The basic element of the common ion pump is a Penning trap. A swirling cloud of electrons produced by an electric discharge are temporarily stored in the anode region of a Penning trap. These electrons ionize incoming gas atoms and molecules. The resultant swirling ions are accelerated to strike a chemically active cathode (usually titanium). On impact the accelerated ions will either become buried within the cathode or sputter cathode material onto the walls of the pump. The freshly sputtered chemically active cathode material acts as a getter that then evacuates the gas by both chemisorption and physisorption resulting in a net pumping action. Inert and lighter gases, such as He and H2 tend not sputter and are absorbed by physisorption. Some fraction of the energetic gas ions (including gas that is not chemically active with the cathode material) can strike the cathode and acquire an electron from the surface neutralizing it as it rebounds. These rebounding energetic neutrals are buried in exposed pump surfaces.

Both the pumping rate and capacity of such capture methods are dependent on the specific gas species being collected and the cathode material absorbing it. Some species, such as carbon monoxide, will chemically bind to the surface of a cathode material. Others, such as hydrogen, will diffuse into the metallic structure. In the former example, the pump rate can drop as the cathode material becomes coated. And, in the latter, the rate remains fixed by the rate at which the hydrogen diffuses.

There are three main types of ion pumps, the conventional or standard diode pump, the noble diode pump and the triode pump.

Ion pumps are commonly used in ultra-high vacuum (UHV) systems, as they can attain ultimate pressures less than 10−11 mbar. In contrast to other common UHV pumps, such as turbomolecular pumps and diffusion pumps, ion pumps have no moving parts and use no oil. They are therefore clean, need little maintenance, and produce no vibrations. These advantages make ion pumps well-suited for use in scanning probe microscopy and other high-precision apparatus.

High Optical Quality Torr Scientific Vacuum Window used on ALBA

Print Friendly, PDF & Email

Located in Barcelona, ALBA is a Synchrotron Light Source producing radiation for experiments in many fields of science, www.cells.es. The quality of the radiation delivered to users depends on the quality of the electron beam and it is therefore necessary to continuously monitor the beam status in a non-invasive way. At ALBA, the visible part of the synchrotron radiation itself is used to measure distribution of electrons in the machine. The used synchrotron radiation is guided far from the machine tunnel up to an “extraction mirror”. In order to maintain the vacuum in the tunnel, the radiation routed up to the mirror is in-vacuum too. A Torr Scientific viewport is used for extracting the light. The goal this year is to measure the transverse beam size by using the interference pattern produced by the synchrotron radiation passing through a double slits aperture. The performance of this kind of experiment requires the preservation of the quality of the light. Torr Scientific therefore provided a special viewport with a flatness better than lambda/10 and a curvature better than 1 arc sec. Accelerator Division Researcher Laura Torino